The Mastersizer 3000+ Lab is an entry level system for particle size analysis by laser diffraction. This system offers robust size measurements from 0.1 to 1000 microns and benefits from the core experience of the Mastersizer Xplorer software.
Features and benefits
The Mastersizer 3000+ Lab provides a cost effective, entry level particle sizing experience which can be upgraded as required.
Particle size range from 0.1 - 1000 µm
Manual dispersion units only
Basic Mastersizer Xplorer software, with updates and bug fixes only
Upgrade option to Mastersizer 3000+ Pro or Ultra
How it works
The Mastersizer 3000+ Lab uses the technique of laser diffraction to measure the size of particles. It does this by measuring the intensity of light scattered as a laser beam passes through a dispersed particulate sample. This data is then analysed to calculate the size of the particles that created the scattering pattern.
A typical system is made up of three main elements:
Optical bench: A dispersed sample passes though the measurement area of the optical bench, where a laser beam illuminates the particles. A series of detectors then accurately measure the intensity of light scattered by the particles within the sample over a wide range of angles
Sample dispersion units (accessories): Sample dispersion is controlled by a range of wet and dry dispersion units. These ensure the particles are delivered to the measurement area of the optical bench at the correct concentration and in a suitable, stable state of dispersion
Instrument software: The basic Mastersizer Xplorer software controls the system during the measurement process and analyzes the scattering data to calculate a particle size distribution
Specification Mastersizer Mastersizer 3000+ Lab
General
Particle size
Suspensions, emulsions, dry powders
Principle
Laser light scattering
Analysis
Mie and Fraunhofer scattering
Data acquisition rate
10 kHz
Typical measurement time
<10 sec
Dimensions (W, D, H)
690mm x 300mm x 450mm
Weight
30 kg
Optics
Red light source
Max. 4mW He-Ne, 632.8nm
Blue light source
None
Lens arrangement
Reverse Fourier (convergent beam)
Effective focal length
300mm
Detector
Arrangement
Log-spaced array
Angular range
0.032 - 60 degrees
Alignment
Automatic
Size
Particle size
0.1 - 1000 µm *
Number of size classes
100 (user adjustable)
Accuracy
Better than 0.6% **
Precision / Repeatability
Better than 0.5% variation *
Reproducibility
Better than 1% variation *
System compliance
Laser safety
Class 1, IEC60825-1:2007 and CFR Chapter I: Sub-chapterJ: Part 1040 (CDRH)
Regulatory testing
RoHS and WEEE compliant CE / FCC compliant Meets requirements of the European Low Voltage directive
System
Power
100/240 v, 50/60 Hz 50W (no dispersion units connected) 200W maximum (2 dispersion units connected)
Humidity
80% maximum for temperatures up to 31°C, decreasing linearly to 50% at 40°C. Non condensing.
Operating temperature (°C)
+5°C to +40°C
Product storage temperature
-20°C to +50°C
Patents
Patents
The Mastersizer optical bench is protected by patents based on applications WO2013038161, WO2013038160 and WO2013038159.
* Sample and sample preparation dependent
** Accuracy defined for the measurement of monomodal latex standards. This specification accounts for the manufacturer’s uncertainty in the latex size. Sample and sample preparation dependent.
This site uses cookies and similar technologies to ensure maximum user experience. By using this site, you confirm your consent to the use of cookies in accordance with this notice regarding this type of files.