Mastersizer 3000+ Pro – Innovative Particle Size Analysis with High Precision | Makrolab LTD
The Mastersizer 3000+ Pro is a mid tier system for particle size analysis by laser diffraction. This system offers robust size measurements from 0.1 to 1000 microns and benefits from a more advanced software experience with Mastersizer Xplorer and automated dispersion options.
Features and benefits Mastersizer 3000+ Pro
The Mastersizer 3000+ Pro generates the highest quality data for applications needing micron scale sizing capabilities and the advanced software options of Mastersizer Xplorer.
Robust and reliable sizing performance in the range from 0.1 - 1000 µm - perfect if you do not need sub-micron or millimeter scale size information
User-friendly software with built in features for automatic measurement setup, method development and live data monitoring. Includes new Data Quality Guidance and SOP Architect features
Flexibility to meet your sample requirements with a range of easily interchanged manual and automated accessories tailored for measurements of suspensions, emulsions and powders
The most connected particle size experience with Smart Manager for live health monitoring to limit downtime
Upgrade option to Mastersizer 3000+ Ultra
How it works Mastersizer 3000+ Pro
The Mastersizer 3000+ Pro uses the technique of laser diffraction to measure the particle size and particle size distribution of materials. It does this by measuring the intensity of light scattered as a laser beam passes through a dispersed particulate sample. This data is then analyzed to calculate the size of the particles that created the scattering pattern. Three main elements enable the system to consistently and reliably deliver the accurate, robust particle sizing data that is central to the effective operation of so many industrial processes.
Optical bench : Mastersizer 3000+ Pro delivers measurements from 0.1 to 1000 µm using a single optical measurement path, making it suitable for an extremely wide range of applications. The dispersed sample passes though the measurement area of the optical bench, where a laser beam illuminates the particles. A series of detectors then accurately measure the intensity of light scattered by the particles within the sample for both red and blue light wavelengths and over a wide range of angles.
Sample dispersion units : Mastersizer 3000+ Pro offers a choice of six wet dispersion accessories to handle an extended range of sample volumes and a wide variety of dispersants. A state-of-the-art dry dispersion system enables rapid and reproducible powder dispersion, even for more fragile materials. Sample dispersion units ensure that particles are delivered to the measurement area of the optical bench at the correct concentration and in a suitable stable state of dispersion to ensure accurate, reproducible measurements.
Mastersizer software : Mastersizer Xplorer software meets the growing demand for easy-to-use instruments that deliver excellent results without constant input from experts. The software controls the system during the measurement process, analyzing scattering data to calculate a particle size distribution. Its intuitive interface guides users through every stage of the process, from selecting a robust method through to routine measurement and result reporting. In-built expert advice on method development, method performance and results quality reduces training requirements and makes particle size analysis simpler and faster.
Specification Mastersizer Mastersizer 3000+ Pro
General
Particle size
Suspensions, emulsions, dry powders
Principle
Laser light scattering
Analysis
Mie and Fraunhofer scattering
Data acquisition rate
10 kHz
Typical measurement time
<10 sec
Dimensions (W, D, H)
690mm x 300mm x 450mm
Weight
30 kg
Optics
Red light source
Max. 4mW He-Ne, 632.8nm
Blue light source
None
Lens arrangement
Reverse Fourier (convergent beam)
Effective focal length
300mm
Detector
Arrangement
Log-spaced array
Angular range
0.032 - 60 degrees
Alignment
Automatic
Size
Particle size
0.1 - 1000 µm *
Number of size classes
100 (user adjustable)
Accuracy
Better than 0.6% **
Precision / Repeatability
Better than 0.5% variation *
Reproducibility
Better than 1% variation *
System compliance
Laser safety
Class 1, IEC60825-1:2007 and CFR Chapter I: Sub-chapterJ: Part 1040 (CDRH)
Regulatory testing
RoHS and WEEE compliant CE / FCC compliant Meets requirements of the European Low Voltage directive
System
Power
100/240 v, 50/60 Hz 50W (no dispersion units connected) 200W maximum (2 dispersion units connected)
Humidity
80% maximum for temperatures up to 31°C, decreasing linearly to 50% at 40°C. Non condensing.
Operating temperature (°C)
+5°C to +40°C
Product storage temperature
-20°C to +50°C
Patents
Patents
The Mastersizer optical bench is protected by patents based on applications WO2013038161, WO2013038160 and WO2013038159.
* Sample and sample preparation dependent
** Accuracy defined for the measurement of monomodal latex standards. This specification accounts for the manufacturer’s uncertainty in the latex size. Sample and sample preparation dependent.