The Mastersizer 3000+ Ultra is our most advanced system for particle size and size distribution by laser diffraction. This model measures the widest size range (0.01 – 3500 um) and supports automation. The digitally enhanced Mastersizer Xplorer comes with a full set of artificial intelligence empowered workflow applications including the new Size Sure measurement mode.
The Mastersizer 3000+ Ultra is the future of particle size analysis.
The Mastersizer 3000+ Ultra is a versatile, compact instrument with state-of-the art software capabilities to meet the analytical and workload demands of the laboratories of present and the future. Combining expert engineering, built-in applications know-how and advanced connectivity, it offers:
Malvern Panalytical's extensive knowledge of laser diffraction particle size analysis enables the Mastersizer 3000+ Ultra to quickly and consistently deliver results you can rely on, with software solutions that empower both measurement novices and particle sizing experts alike.
The Mastersizer 3000+ Ultra uses the technique of laser diffraction to measure the particle size and particle size distribution of materials. It does this by measuring the intensity of light scattered as a laser beam passes through a dispersed particulate sample. This data is then analyzed to calculate the size of the particles that created the scattering pattern. Three main elements enable the system to consistently and reliably deliver the accurate, robust particle sizing data that is central to the effective operation of so many industrial processes.
Optical bench : Mastersizer 3000+ Ultra delivers measurements from 10 nm to 3.5 mm using a single optical measurement path, making it suitable for an extremely wide range of applications. The dispersed sample passes though the measurement area of the optical bench, where a laser beam illuminates the particles. A series of detectors then accurately measure the intensity of light scattered by the particles within the sample for both red and blue light wavelengths and over a wide range of angles.
Sample dispersion units : Mastersizer 3000+ Ultra offers a choice of six wet dispersion accessories to handle an extended range of sample volumes and a wide variety of dispersants. A state-of-the-art dry dispersion system enables rapid and reproducible powder dispersion, even for more fragile materials. Sample dispersion units ensure that particles are delivered to the measurement area of the optical bench at the correct concentration and in a suitable stable state of dispersion to ensure accurate, reproducible measurements.
Mastersizer software : Mastersizer Xplorer software meets the growing demand for easy-to-use instruments that deliver excellent results without constant input from experts. The software controls the system during the measurement process, analyzing scattering data to calculate a particle size distribution. Its intuitive interface guides users through every stage of the process, from selecting a robust method through to routine measurement and result reporting. In-built expert advice on method development, method performance and results quality reduces training requirements and makes particle size analysis simpler and faster.
General
Particle size | Suspensions, emulsions, dry powders |
Principle | Laser light scattering |
Analysis | Mie and Fraunhofer scattering |
Data acquisition rate | 10 kHz |
Typical measurement time | <10 sec |
Dimensions (W, D, H) | 690mm x 300mm x 450mm |
Weight | 30 kg |
Red light source | Max. 4mW He-Ne, 632.8nm |
Blue light source | Max. 10mW LED, 470nm |
Lens arrangement | Reverse Fourier (convergent beam) |
Effective focal length | 300mm |
Arrangement | Log-spaced array |
Angular range | 0.015 - 144 degrees |
Alignment | Automatic |
Particle size | 0.01 - 3500 µm * |
Number of size classes | 100 (user adjustable) |
Accuracy | Better than 0.6% ** |
Precision / Repeatability | Better than 0.5% variation * |
Reproducibility | Better than 1% variation * |
21 CFR Part 11 | Enables an operating mode that assists with ER/ES compliance |
Laser safety | Class 1, IEC60825-1:2007 and CFR Chapter I: Sub-chapterJ: Part 1040 (CDRH) |
Regulatory testing | RoHS and WEEE compliant CE / FCC compliant Meets requirements of the European Low Voltage directive |
Power | 100/240 v, 50/60 Hz 50W (no dispersion units connected) 200W maximum (2 dispersion units connected) |
Humidity | 80% maximum for temperatures up to 31°C, decreasing linearly to 50% at 40°C. Non condensing. |
Operating temperature (°C) | +5°C to +40°C |
Product storage temperature | -20°C to +50°C |
Patents | The Mastersizer optical bench is protected by patents based on applications WO2013038161, WO2013038160 and WO2013038159. |